I am doing a comparison of a CIRS phantom in clarius using H-scan
While in verasonics for the same phantom the map is like this
In clarius C3 (3.5Mhz)
In clarius L15 (10Mhz) is like this:

I realized while in verasonics the spectrum of the RF data is centered like this

In clarius C3 spectrum
In clarius the L15 spectrum (after passing a high pass filter at 0.5Mhz - Because I realized there is a DC component) is like this:
So the fitting of the RF data for L15 is a bit weird. I wanted to know if are you using harmonics or why do I have signal after 11 Mhz (raw spectrum). what would be a recommended bandpass filter then?
As an extra question, I am using RF data to reconstruct my Bmode. I wanted to know if I use IQ data this IQ data has better resolution and if I can modulate this IQ into a RF data to have an improved spectra.
Thanks in advance.
Hello,
I am doing a comparison of a CIRS phantom in clarius using H-scan
While in verasonics for the same phantom the map is like this
In clarius C3 (3.5Mhz)
In clarius L15 (10Mhz) is like this:

I realized while in verasonics the spectrum of the RF data is centered like this

In clarius C3 spectrum
In clarius the L15 spectrum (after passing a high pass filter at 0.5Mhz - Because I realized there is a DC component) is like this:
So the fitting of the RF data for L15 is a bit weird. I wanted to know if are you using harmonics or why do I have signal after 11 Mhz (raw spectrum). what would be a recommended bandpass filter then?
As an extra question, I am using RF data to reconstruct my Bmode. I wanted to know if I use IQ data this IQ data has better resolution and if I can modulate this IQ into a RF data to have an improved spectra.
Thanks in advance.